beautypg.com

Installation – KROHNE OPTITEMP TR 100-TC 100 EN User Manual

Page 14

background image

3

INSTALLATION

14

OPTITEMP TR/TC 100

www.krohne.com

10/2009 - 4000630002 - MA OPTITEMP TR/TC 100 R02 en

Inserting the measuring insert into the thermometer assembly

• Open the cover of the connection head on the thermometer assembly and ensure that the

gasket does not get damaged or lost (contact the local representative for new gaskets).

• Slide the measuring insert completely into the thermowell until the washer disc on the

measuring insert almost touches the floor of the connection head.

• Tighten the two M4 screws on the measuring insert by hand.
• Close the cover of the connection head of the enclosure.

.book Page 14 Thursday, October 8, 2009 10:26 AM