beautypg.com

Installation, 1 pressure and temperature ranges – KROHNE OPTIWAVE 7300C Quickstart EN User Manual

Page 9

background image

INSTALLATION

2

9

OPTIWAVE 7300 C

www.krohne.com

03/2014 - 4000050005 - QS OPTIWAVE 7300 R08 en

2.7 How to prepare the tank before you install the device

2.7.1 Pressure and temperature ranges

CAUTION!
To avoid measuring errors and device malfunction, obey these precautions.

Figure 2-8: Pressure and temperature ranges

1 Flange temperature

FKM/FPM gasket: -40...+200°C / -40...+390°F; Kalrez

®

6375 gasket: -20...+200°C / -4...+390°F;

EPDM gasket: -50...+150°C / -58...+300°F
Depends on the antenna type. Refer to the table that follows.
Ex devices: see supplementary operating instructions

2 Ambient temperature for operation of the display

-20...+60°C / -4...+140°F
If the ambient temperature is not between these limits, the display screen switches off automatically

3 Ambient temperature

Non-Ex devices: -40...+80°C / -40...+175°F
Ex devices: see supplementary operating instructions

4 Process pressure

Depends on the antenna type. Refer to the table that follows.

WARNING!
The process connection temperature range must agree with the temperature limits of the
gasket material. The operating pressure range is subject to the process connection used and the
flange temperature.

Antenna type

Maximum process connection

temperature

Maximum operating pressure

[°C]

[°F]

barg

psig

PP Drop

+100

+210

16

232

PTFE Drop

+150

+300

40

580

Hygienic

+150

+300

10

145

Horn / Sheet metal horn

+150 (+200)

1

+300 (+390)

1

40 (100)

2

580 (1450)

2

1 Standard max. process connection temperature: +150°C / +300°F. Optional max. process temperature: +200°C /

+390°F.

2 Standard max. operating pressure: 40 barg / 580 psig. Optional max. operating pressure: 100 barg / 1450 psig.