Technical data – KROHNE OPTIWAVE 7300C Quickstart EN User Manual
Page 27
TECHNICAL DATA
4
27
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000050005 - QS OPTIWAVE 7300 R08 en
Installation conditions
Process connection size
The nominal diameter (DN) should be equal to or larger than the antenna diameter.
If the nominal diameter (DN) is smaller than the antenna, either:
– provide the means to adapt the device to a larger process connection on the tank
(for example, a plate with a slot), or
– use the same process connection, but remove the antenna from the device before
installation and fit it from inside the tank.
Process connection position
Make sure that there are not any obstructions directly below the process
connection for the device.
Dimensions and weights
Refer to "Technical data: Dimensions and weights" in the Handbook.
Materials
Housing
Standard: Polyester-coated aluminium
Option: Stainless steel (1.4404 / 316L)
2
Wetted parts, including antenna
Standard for Horn / Sheet metal horn antenna: Stainless steel (1.4404 / 316L)
Option for Horn antenna: Hastelloy
®
C-22 (2.4602)
3
Standard for Drop antenna: PTFE; PP
Option for Drop antenna: PP or PTFE flange plate protection
Hygienic antenna: PEEK – this material agrees with FDA regulations
Process connection
Standard for Horn, Sheet metal horn and Drop antennas:
Stainless steel (1.4404 / 316L) – a PP or PTFE flange plate protection option is also
available for the Drop antenna
Standard for Hygienic antenna: PEEK
Option: Hastelloy
®
C-22 (2.4602) – for Horn antennas only
Gaskets (and o-rings for the
sealed antenna extension option)
Hygienic antenna:
Hygienic antenna:
Hygienic antenna:
Hygienic antenna:
BioControl
®
: FKM/FPM (-20…+150°C / -4…+300°F); EPDM (-20°C…+150°C /
-4…+300°F)
SMS, Tri-Clamp
®
, DIN 11851: without
4
PTFE Drop antenna:
PTFE Drop antenna:
PTFE Drop antenna:
PTFE Drop antenna:
FKM/FPM (-40…+150°C / -40…+300°F); Kalrez
®
6375 (-20…+150°C / -4…+300°F);
EPDM (-50°C…+150°C / -58…+300°F)
5
PP Drop antenna:
PP Drop antenna:
PP Drop antenna:
PP Drop antenna:
FKM/FPM (-40…+100°C / -40…+210°F); Kalrez
®
6375 (-20…+100°C / -4…+210°F);
EPDM (-40°C…+100°C / -40…+210°F)
5
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
FKM/FPM (-40…+200°C / -40…+390°F); Kalrez
®
6375 (-20…+200°C / -4…+390°F);
EPDM (-50°C…+150°C / -58…+300°F)
5
Feedthrough
Standard: PEI (-50...+200°C / -58...+390°F – max. range. The feedthrough
temperature limits must agree with the temperature limits of the gasket material
and antenna type. If the distance piece option is not attached, the maximum
temperature is 150°C / 300°F.)
Option: Metaglas
®
(-30...+200°C / -22...+390°F – max. range. The feedthrough
temperature limits must agree with the temperature limits of the gasket material
and antenna type. If the distance piece option is not attached, the maximum
temperature is 150°C / 300°F.)
6
Weather protection (Option)
Stainless steel (1.4301 / 304)
Process connections
Thread
G 1½ (ISO 228); 1½ NPT (ASME B1.20.1)