Installation – KROHNE OPTIWAVE 7300C Quickstart EN User Manual
Page 15
INSTALLATION
2
15
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000050005 - QS OPTIWAVE 7300 R08 en
• Make sure the tank connection is level.
• Make sure that you use the applicable gasket for the connection dimensions and the process.
• Align the gasket correctly.
• Lower the antenna carefully into the tank.
• Turn the threaded connection on the housing to attach the device to the process connection.
• Tighten the connection.
i
Refer to local rules and regulations for the correct torque to apply to the connection.
For more data, refer to the handbook.
2.9.3 How to install a device with a hygienic connection
BioControl
®
Equipment needed:
• Device
• Gasket
• Flange bolts (not supplied)
• Wrench (not supplied)
How to attach a device with a Biocontrol
®
connection
• Make sure that the flange on the nozzle is level.
• Make sure that you use the applicable gasket for the flange dimensions and the process.
• Align the gasket correctly on the flange facing of the nozzle.
• Lower the antenna carefully into the tank.
• Tighten the flange bolts.
i
Refer to local rules and regulations for the correct torque to apply to the bolts.
INFORMATION!
To make the cleaning of the antenna easier, attach the device to a short socket.
Figure 2-17: BioControl
®
connection
1 BioControl
®
connection on the tank
2 Flange bolts