Technical data – KROHNE OPTIWAVE 7300C Quickstart EN User Manual
Page 26
4
TECHNICAL DATA
26
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000050005 - QS OPTIWAVE 7300 R08 en
Measuring accuracy
Resolution
1 mm / 0.04¨
Repeatability
±1 mm / ±0.04¨
Accuracy
±3 mm / ±0.12¨, when distance < 10 m / 33 ft;
±0.03% of measured distance, when distance > 10 m / 33 ft
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Temperature
+20°C ±5°C / +70°F ±10°F
Pressure
1013 mbara ±20 mbar / 14.69 psia ±0.29 psi
Relative air humidity
60% ±15%
Target
Metal plate in an anechoic chamber
Operating conditions
Temperature
Temperature
Temperature
Temperature
Ambient temperature
-40…+80°C / -40…+175°F (according to the temperature limits of the gasket
material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Storage temperature
-40…+85°C / -40…+185°F
Process connection temperature
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -50…+150°C / -58…+300°F
Option: -50…+200°C / -58…+390°F
(the process connection temperature must agree with the temperature limits of the
gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-50…+150°C / -58…+300°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-40…+100°C / -40…+210°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
-20…+150°C / -4…+300°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Pressure
Pressure
Pressure
Pressure
Operating pressure
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-1…16 barg / -14.5…232 psig;
subject to process connection used and flange temperature
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-1…40 barg / -14.5…580 psig;
subject to process connection used and flange temperature
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
-1…10 barg / -14.5…145 psig;
subject to process connection used and flange temperature
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -1…40 barg / -14.5…580 psig;
Option: -1…100 barg / -14.5…1450 psig;
subject to process connection used and flange temperature
Other conditions
Other conditions
Other conditions
Other conditions
Dielectric constant (ε
r
)
≥1.5
Vibration resistance
IEC 60068-2-6 and EN 50178 (10...57 Hz: 0.075 mm / 57...150 Hz:1g)
Ingress protection
IP 66/67 equivalent to NEMA type 4X (housing) and type 6P (antenna)