Specification, 12 specification, 52 specification – KANOMAX 6501 series Climomaster User Manual
Page 53

52 Specification
§ 12
Specification
Product
CLIMOMASTER Air Velocity Meter
Model No. of the Main Body
6501
Model No. of the Probe
6531-21 6541
-21
6561
-21
6542
-21
6551
-21
6552
-21
6533-21
6543
-21
Measuring Object
Clean air flow
Air Velocity
[m/s]
Measuring Range
0.01 to 30.0, 0.01 to 50.0 (6561-21 only)
0.01 to 5.00
Resolution
0 to 9.99: 0.01
10 to 30: 0.1
30 to 50: 0.1 (6561-21 only)
0.01
Accuracy
±2% of reading or ±0.015 (±3FPM),
whichever is greater
0.01 to 0.99:±0.02
0.99 to 5.00:
±2% of reading or
±0.015 (±3FPM),
whichever is greater
Response Time
[sec.]
Approx. 1
Approx. 4
Approx. 7
Approx. 7
(at 1m/s, 90% response time)
Air
Temperature
[
o
C]
Measuring Range
-20.0 to 70.0
-
-20.0 to 70.0
Resolution
0.1
-
0.1
Accuracy
±0.5
-
±0.5
Response Time
[sec.]
Approx. 30
(at 1m/s, 90% response time)
-
Approx. 30
(at 1m/s, 90%
response time)
Humidity
*1
[%RH]
Measuring Range 2.0 to 98.0
-
2.0 to 98.0
-
Resolution
0.1
-
0.1
-
Accuracy
2 to 80:
±2.0
80 to 98:
±3.0
-
2 to 80:
±2.0
80 to 98:
±3.0
-
Response Time
[sec.]
Approx.
15
-
Approx.
15
-
Pressure
※
2
[kPa]
Measuring Range
-5.00~+5.00
Resolution
0.01
Accuracy
±3% of reading or ±0.01
Response Time
[sec.]
Approx. 1
Temperature Compensation
[m/s]
±5% of reading or ±0.015 (±3FPM), whichever is greater
(in the temperature range of 5 to 60.0
o
C)
Functions
Readings Hold, Max Hold, Time Constant (1, 5, 10 sec.), Battery Level
Indicator (5 Levels),
Unit selection (Air Velocity: m/s or FPM, Flow Rate: m
3
/min, m
3
/h,
ft
3
/min, or ft
3
/h, Temperature
*1
:
o
C or
o
F, Humidity
*2
: %RH, Static
Pressure
*3
: kPa or Pa)
Max., Min., and Average Calculation, Sampling Time: 1 to 999 sec.
No. of trails: 1 to 9999, Store up to 25 Duct Sizes (Square/Round, Size
Range: 1 to 9999mm or 0.1 to 999.9inch(diameter, height, width))
Max.Data Storage: 20000
Output
Digital Output: USB (when connected to a printer, automatically switch
to RS-232C, Baud Rate 4800, 9600, 19200 and 38400 bps)
Analog Output
*3
: DC 0 to 1V (select one from Air Velocity,
Temperature
*1
, Humidity
*2
and Pressure
*3
)
Power
6 × AA Batteries , AC Adaptor
*4
: AC 100 to 240V (50/60Hz)
Battery Life
Approximately 10 continuous hours (at air velocity 5m/s, 20
o
C, with
alkaline batteries)
Operating Environment
Main Body: 5 to 40
o
C (41 to 104
o
F) with no visible condensation
Probe: -20 to 70
o
C (-4 to 158
o
F) with no visible condensation
Storage Environment
-10 to 50
o
C (14 to 122
o
F) with no visible condensation
Weight
Approximately 400g (batteries included)
Accessories
Carrying case, Operation manual, 6×AA Batteries, Probe cable (2m),
USB cable
Options
Spare probe, Analog output, Pressure sensor, Extension rod, Printer,
Printer cable, Software (for Windows), AC adaptor, Hands-free Case
*
1: Air temperature is only available on MODEL6531/6541/6542/6533/6543/6561
*
2: Humidity is only available on MODEL6531/6533
*
3: Pressure is only available on MODEL6501-B0/-C0 , Analog Output is only available on MODEL6501-A0/-C0
*
4: Optional.