0 0.00h, Pressure, Zero adjustment procedure – KANOMAX 6501 series Climomaster User Manual
Page 16: Connecting the pressure tube

15
Pressure
●
Do not expose the pressure sensor to more than 75kPa of pressure.
Excess pressure may cause serious damage to the sensor.
●
The operating temperature must be between 5
o
C and 40
o
C or 41
o
F
and 104
o
F when measuring pressure. The instrument may not
operate properly outside of this temperature range.
●
Make sure to perform a zero adjustment before measuring
pressure. When performing a zero adjustment, leave both pressure
ports (+) and (-) open.
●
The zero adjustment range is – 0.5 to 0.5 kPa. If the reading is out
of this range, an error message will be displayed.
■
Zero Adjustment Procedure
Press
to enter the Main Menu.
Use
to select “8. PRESSURE ZERO”
and press
.
Use
to select “1. ZERO ADJUST”
and press
.
Use
to select “YES”.
Press
to display 0.00 kPa.
Use
to select “2. SAVE INFO”.
Press
. This will complete the zero
adjustment and you are back in Main
Menu.
If you press
before you save, you will go back to the Main Menu
without completing the zero adjustment.
■
Connecting the Pressure Tube
Connect the pressure tube to the (+) or (-)
port as you see on the left.
Connect/insert the other end of the tube to the
measuring hole of the duct. If the pressure
that you are going to measure is positive,
connect the tube to (+) port. If it is negative,
connect the tube to (-) port.
aaaa
PRESSURE
a
ZERO
1.ZERO
a
ADJUST
aaa
YES
2
.SAVE
a
INFO
0
0.00h
aaaa
PRESSURE
a
ZERO
1.ZERO
a
ADJUST
aaaY
ES
2.SAVE
a
INFO
0
0.00h
aaaa
PRESSURE
a
ZERO
1
.ZERO
a
ADJUST
aaa
YES
2.SAVE
a
INFO
-0.00h
MENU
aa
1.NORMAL
aaaaaa
2.DUCT
a
TYPE
aaaaaa
3.CALCULATION
aaaaaa
4.FLOW
a
RATE
aaaaaa
5.DATA
a
OUTPUT
aaaaaa
6.DATA
a
CLEAR
aaaaaa
7.UTILITY
aaaaaa8
.PRESSURE
a
ZERO
*
In order to take
an accurate
measurement,
make sure that the
tube is properly
secured without
any leakage, or
bends.
■
Pressure sensor is
only available on
MODEL6501-B0/-C0.