3. Display Description and Operation Procedure
17
[CONTINUOUS MODE]
Start/Stop measuring manually.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.2 CONTINU):
INTERVAL :
N/A
SAMPLE T.
: N/A
CYCLES
: N/A
A continuous measurement is performed until you manually stop.
Measurement Flow
[INTERVAL MODE]
Configure INTERVAL, SAMPLE T and CYCLES, and repeat measuring.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.3 INTERVAL):
INTERVAL :
00:01:00
SAMPLE T.
: 00:00:15
CYCLES
: 20
A total of 20 samples are performed at one minute intervals.
Note: Sampling time is 15 seconds and standby time is 45 seconds.
Measurement Flow
If the time period between two samplings is 10 seconds or longer, the pump will be turned OFF
after a sampling is finished. Then the pump will be turned ON again 10 seconds before the next
sampling starts.
Sampling Time
Start
Measurement
Finish
Measurement
Wait Time
STOP
Interval
Sampling Time
(15sec)
The 1
st
Measurement
The 2
nd
Measurement
The 20
th
Measurement
Total Time: 19min 15sec
Start
Measurement
Finish
Measurement
Wait Time
Interval
Sampling Time
(15sec)
Sampling Time
(15sec)