2. Getting Started
4
After pressing the POWER switch on the front of the instrument, you will hear a beep
sound and the LED will light up. If the power is supplied by the AC line, the LED is
green and if supplied by battery, the LED is red.
Several seconds later, the initial screen appears automatically. Refer to 3. Display
Description and Operation Procedure for the details about the operation screen.
2.2 Attaching Standard Inlet
Upon measuring, attach the standard inlet following the procedure below.
Remove rubber cap
Prepare standard inlet
Tighten screw
Done!
Remove the rubber cap that is mounted on the inlet (under the
handle) on the upper side of the main unit.
Put the screw supplied with the instrument on the inlet. (See the
photo on the right.)
Attach the standard inlet supplied with the instrument to the inlet.
The knurling side of the port goes into the inlet.
Hold the top of the inlet port down slightly when tightening a
screw counterclockwise.
If the screw is tightly fit into the hole on the handle, the inlet is
mounted properly.
To take the inlet out, reverse the procedure.
Knurling Side
Standard Inlet
Inlet
Screw