Main specifications – KANOMAX 3522 Piezobalace Dust Monitor User Manual
Page 27
22
9. Main Specifications
Product Name
Piezobalance Dust Monitor
Model
3521
Measuring Object
Mass concentration of suspended particulate matter in indoor or
outdoor air.
Operating Temperature Range
-10°C to 40°C (14°F to 104°F)
Storage Temperature Range
-30°C to 60°C (-22°F to 140°F)
Measuring Range
0.01 to 5mg/m
3
Measuring Particle Size
Φ 10 µm (below 98%) or Φ 4 µm (below 50%)
(depends on which impactor nozzle is used)
Particle Sizing Method
Impactor (Inertial Impaction)
Measuring Time
24s, 120s, 1-60 min (Preset by the menu)
Measuring Accuracy
±10% of the calibration particles
Resolution 0.01mg/m
3
Suction Flow
1 L/min
Cleaning System
Manual cleaning of the sensor
Communication Protocol
Standard RS232C
Communication Baud Rate
4800bps, 9600bps, 19200bps (Preset by the menu)
Maximum Data Records
500 records
Weight
Approx. 1.8 kg (4 lbs.)
Dimensions
65Ч180Ч150 mm (2.6Ч7.1Ч5.9 inches) (W×L×H)
Power Source
1) AC/DC adapter (input: AC100-240V)
2) Built-in Ni-MH battery (9.6V-1.5AH),
Continuous operation: approx. 4.5h with 5-8h charging time
Standard Accessories
1) Cleaning Device ……………… 3
2) Detergent/Purified Water ……………… 30cc bottle (1 ea.)
3) Strap ………………
1
4) AC Adapter
………………
1
5) Φ4µm impactor nozzle ……………… 1
6) Communication Cable ………………
1
7) Communication Software CD-ROM …… 1
8) Operation Manual
………………
1
9) Carrying Case
………………
1
Options
10) Printer ………………
1
11) Printer AC/DC Adapter ………………
1
12) Printer Cable
……………… 1
* Certain test functions required in China are not included