Multi-Parameter
Water Quality TROLL
®
DESCRIPTION
The instrument’s Viton® o-rings help to insure a water-tight seal. If the o-rings
appear worn, dried, nicked, stretched, cracked, etc., they should be replaced. If in
any doubt, it is best to err on the side of safety and replace them.
The kit includes dust caps, sensor port plugs, lubricant, and o-rings in the following
sizes (small to large):
0.301” I.D., 0.441" O.D. (#011), for sensors and sensor port plugs (qty: 8)
0.364" I.D., 0.504" O.D. (#012), for back end cable connector (qty: 2)
1.176" I.D., 1.316" O.D. (#025), for top of battery compartment (qty: 2)
1.489” I.D., 1.629" O.D. (#029), 2 at bottom of battery compartment, 1 on nose cone
(qty: 4)
38 mm I.D., 42 mm O.D., 2 mm thick, for front end of body, seal to restrictor and Cal
Cup (qty: 2)
REPLACEMENT PROCEDURE
1. Use your fingers or a non-metallic tool to remove the worn o-rings from the grooves.
2. Clean all mating surfaces with a swab or lint-free cloth.
3. Using your fingers, apply lubricant evenly to new o-rings.
4. Place the o-rings in the grooves. Remove excess lubricant.
GUIDELINES & PRECAUTIONS
•
Be careful not to transfer grease from your fingers to the pins in the cable
connector, or to the sensor connectors and sensor port connectors. Any
lubricant in these areas should be removed with a clean cotton swab.
•
Always use a dust cap to protect the o-ring and connector when cable is not
attached.
•
The kit includes sensor port plugs for deployments with fewer than 4
removable sensors installed.
MP TROLL 9500
MAINTENANCE KIT
Catalog No. 95100
0095102 10/05
Every sensor port must be
occupied by a sensor or plug
before submersion.
The o-rings labeled D and E are
nearly identical in size. However,
E is slightly thicker than D
A
B
C
D
E
Actual Size
A
B
C
D
E
Sensor port
plug
A
C
D
Dust cap
B
D
E