beautypg.com

In-Situ Nose Replacement Kit for the optical RDO PRO Probe User Manual

Instruction sheet, Nose replacement kit for the rdo, Pro and rdo

background image

Instruction Sheet

Nose Replacement Kit for the RDO

®

PRO and RDO

®

PRO-X Probe

Part Number 0080820

Kit Contents

1

Replacement Nose Cone

2

Installed O-ring

3

O-ring for probe body

4

Lubricant

5

Lens cloth

Replacement Procedure

1. Remove the nose cone and existing O-ring from the

probe body.

1

O-ring

2

Nose Cone

2. Place a small amount of lubricant on your fingertip.

Coat the entire O-ring with lubricant. Do not allow any
dust or debris to get on the O-ring.

3. Place the new O-ring on the probe body.

4. Use the lens cloth to remove any lubricant from the

probe and cap.

5. Place the nose cone on the probe body.

Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.

©

2013. All rights reserved.

11/2013 | Rev. 002 | 0080822

Table of contents