Multi-Parameter

Water Quality TROLL

®

A

NOSE CONE STIRRER

Catalog No. 42210

APPLICATION
Battery-powered magnetic stirrer provides continuous agitation of sample during
calibration or deployment in slow-moving or stagnant water. Attaches to the front
end of the MP TROLL in place of the nose cone, or to the Calibration Cup. Elimi-
nates potential sensor cross-contamination and increases the rate of flow through
the restrictor. Guard plate retains stir bar in fast-moving water and protects it from
debris.

PHYSICAL DESCRIPTION
Wetted materials

Delrin

®

, Teflon

®

Dimensions

4.6 cm (1.8 in) dia. x 19 cm (7.5 in) long

Weight

0.575 kg (1.3 lb), with batteries

Operating voltage

1.5 - 9.0 V, nominal 3 V

Motor speed

950 rpm at free load, 350-430 operating rpm

Stall torque

32 g-cm

Current

0.052 amps

Output

0.073 watts

Stir bar

Teflon-coated, magnetic, 7.9 mm (5/16 in) dia. x
25.4 mm (1 in) long with moldring

Battery type

2 standard alkaline D-cells

Est. battery life

3 months continuous operation

Pressure rating

150 psi

CONNECTIONS
The threads at the stir bar end attach to the restrictor at the front end of the MP
TROLL (nose cone removed), or to the Cal Cup.

GUIDELINES AND PRECAUTIONS

To activate the stirrer, tighten the end cap about ¼ turn. This starts the motor;
the stir bar will begin to spin. To stop the motor, back off the cap until the stir bar
stops spinning.

Replace batteries with 2 standard alkaline D-cells, – toward the end cap,
+ toward the stir bar compartment.

When attaching to the restrictor or Cal Cup, be careful not to over-tighten.

The stir bar guard may be removed for cleaning when necessary. Insert the
sensor removal tool into the center rings and pop the guard out.

Inspect o-rings regularly; replace if they show evidence of wear or damage (o-
rings & lubricant available from In-Situ Inc. or your distributor).

0042215 rev. 001 09/05

o-rings

o-ring

Stirrer

Stir bar compartment

Battery
compartment

Motor compartment
(sealed)

End cap

OFF

ON

A

Body

Restrictor

Stirrer

Cal Cup

Body

Stirrer