beautypg.com

In-Situ Sensor Summary Sheet for the TROLL 9500 User Manual

9500 multiparameter instrument, Troll, 9500 sensor summary

background image

TROLL

®

9500 Sensor Summary

Innovations in

Water Monitoring

TROLL

®

9500 Multiparameter Instrument

TROLL 9500 General

Operating temp.

-5 to 50° C (23 to 122° F)

Storage temp.

-40 to 65° C (-40 to 140° F)

Dimensions & weight

4.7 cm (1.85 in) OD x 55.25 cm (21.75 in). With twist-lock hanger: 56.52 cm (22.25 in). Restrictor: 8.9 cm (3.5 in) OD x 21 cm (8.25 in) long;

1.9 kg (4.2 lbs)

Wetted materials

PVC, 316L stainless steel, titanium, Acetal, Viton

®

, nylon. Cable: Tefzel

®

or polyurethane

Water tightness rating

IP68 with all sensors and cable attached. Battery compartment: IP67 without the battery cover or cable attached

Output options

RS485/RS232; SDI-12 (optional with SDI-12 adapter); ASCII streaming mode or binary command

Power

External: 9-16 VDC (optional). Internal: 2 user-replaceable D batteries (use either alkaline or matched pair of Saft lithium batteries)

TROLL 9500 Logging

Data logging

Professional and LTS models only: 16 programmable tests (defined, scheduled to run, or stored). Logging modes: Linear, Linear Average, Event

Memory

4 MB (222,000 data records

1

)

Standard Sensors

Accuracy

Range

Depth Rating

Response Time (T90)

5

Methodology

Barometric pressure

±0.3% FS

16.5 psia

Meets highest rating

<30 sec per 30 m (100 ft)

of cable

Silicon strain gauge

Level, Depth, Pressure

±0.1% FS or better

Sensor accuracy: -5 to 50° C

15, 30, 100, or 300 psi

Non-vented

30 psia: 10.90 m (35.76 ft)

100 psia: 60.11 m (197.2 ft)

300 psia: 200.7 m (658.6 ft)

Vented

15 psig: 10.55 m (34.61 ft)

30 psig: 21.10 m (69.21 ft)

100 psig: 70.32 m (230.7 ft)

300 psig: 211.0 m (692.1 ft)

Instantaneous in thermal

equilibrium

Silicon strain gauge

(non-vented or

vented)

Conductivity

Low: ±0.5% or 2 µS/cm

High: ±0.5% + 2 µS/cm

Low: 5 to 20,000 µS/cm

High

2

: 150 to 112,000 µS/cm

Low: Meets highest rating

High: Meets highest rating

Low: Instantaneous

High: Instantaneous

Std. Methods 2510,

EPA 120.1

Std. Methods 2510,

EPA 120.1

Dissolved oxygen

RDO

®

sensor

3

Clark cell electrode

±0.1 mg/L

±0.2 mg/L

±10% of reading

±0.2 mg/L

0 to 8 mg/L

8 to 20 mg/L

20 to 50 mg/L

0 to 20 mg/L; 0 to 200%

saturation

Meets highest rating

246 m (807 ft)

T90: <45 sec. T95: <60 sec.

T90: <45 sec. T95: <60 sec.

T90: <45 sec. T95: <60 sec.

1-mil membrane: 1-2 min

@ 25° C

2-mil membrane: 90 sec

to 3 min

EPA-approved In-Situ

Methods

4

1002-8-

2009, 1003-8-2009,

1004-8-2009

Std. Methods 4500-O G,

EPA 360.1

pH (single) or

pH/ORP (combo)

6

pH: ±0.1 pH unit

ORP: ±5.0 mV

pH: 0 to 12 pH units

ORP: ±1400 mV

pH: 211 m (692 ft)

pH/ORP: 211 m (692 ft)

pH: <15 sec, pH 7 to pH 4

ORP: <15 sec

pH: Std. Methods 4500-

H

+

, EPA 150.2

ORP: Std. Methods

2580

Temperature

±0.1° C

-5 to 50° C (23 to 122° F)

Meets highest rating

< 30 sec

EPA 170.1

Extended Parameter (XP) Sensors

Ammonium (NH

4

+

)

±10%

0.14 to 14,000 ppm N

14 m (46 ft)

T98: <60 sec, 1.4 to 14

ppm N

Std. Methods 4500-NH

3

D, EPA 350.3

Chloride (Cl

-

)

±15%

0.35 to 35,500 ppm Cl

70 m (231 ft)

T98: <60 sec, 3.54 to 35.45

ppm Cl

Std. Methods 4500-

Cl

-

D

Nitrate (NO

3

-

)

±10%

0.14 to 14,000 ppm N

14 m (46 ft)

T98: <60 sec, 1.4 to 14

ppm N

Std. Methods 4500-

NO

3

D

Turbidity

±5% or 2 NTU/FNU

0 to 2,000 NTU/FNU

105 m (346 ft)

Instantaneous (5 sec for first

reading)

ISO 7027

Warranty

TROLL 9500 and all sensors (excluding RDO & ISE sensors) come with a 1-year warranty. RDO Sensor: 3-year warranty. ISE sensors: 90-day

warranty. RuggedCable

®

System: 2-year warranty.

Notes

1

A single data record includes time stamp, temperature, RDO, pH, and conductivity logged in Linear or Linear Average mode.

2

Full operating

range: 70 to 200,000 µS/cm.

3

Full operating range: 0 to 50 mg/L.

4

EPA-approved under the Alternate Test Procedure process.

5

Time required for a

sensor to reach 90% stable.

6

pH sensor and pH/ORP sensor temperature range 0 to 50° C.