In-Situ RDO Sensor for TROLL 9500: Installation, Calibration, and Care (purchased after 09/2008) User Manual

Optical dissolved oxygen sensor

background image

Instruction Sheet

TROLL

®

9500 Water

Quality Instrument

RDO

®

Optical Dissolved Oxygen Sensor

Cable Connect 0085070
Direct Connect 0084310

DESCRIPTION
The In-Situ RDO

®

optical dissolved oxygen sensor measures dissolved

oxygen (DO) using the principle of “dynamic luminescence quenching.”
Certain molecules, called “lumiphores,” fluoresce when excited by light of
a specific wavelength. Oxygen molecules act to quench this fluorescence.
The lumiphores in the sensor are embedded in a gas-permeable sensing
foil in a replaceable cap.

The sensor optics include a lens, blue LED and filter, red LED and filter,
and a photodetector. When the blue LED emits light, the sensing foil emits
red photons; the presence of oxygen in the foil causes a reduction in red
light detected by the photodiode. The phase difference between the blue
excitation light and the returned red light is measured, and the result is
used to compute DO.

This method measures the "phase" (or delay) of the returned signal, and is
thus based on the “lifetime” rather than the “intensity” of the luminescence.

REquIRED MaTERIalS
To use the RDO sensor, you will need the following items:

TROLL

®

9500 instrument

PC or RuggedReader

®

handheld PC

Most recent Win-Situ

®

4 or Pocket-Situ 4 software

1

RDO sensor package

Alkaline battery package (included with the TROLL 9500 instrument)

Communication cable

Most recent version of TROLL 9500 firmware

TROLL

®

Com cable

1 Make sure to use the most recent version of Win-Situ software. Free updates are available at
www.in-situ.com. You must have the most recent software version to allow firmware updates.

uNPaCk ThE RDO CablE CONNECT SENSOR
The RDO cable connect sensor arrives installed in a 3-part adapter/restrictor/
nose cone assembly with the sensor cap in a black canister. Remove the
black restrictor and nose cone from the sensor and adapter assembly.

uNPaCk ThE RDO DIRECT CONNECT SENSOR
The RDO direct connect sensor ships in a box as shown below:

INSTallINg ThE RDO SENSOR CaP

Remove the red protective dust cap from the sensor and

1.

save it for later use.

Protective dust cap

Sensor cap

Use the supplied lens wipe to clean the lens on the sensor, if necessary.

2.

Remove the RDO sensor cap from its shipping sleeve.

3.

Align the arrow on the cap with the index mark on the sensor and

4.

firmly press the cap onto the sensor, without twisting, until it seals over
the probe body.

Keep the cap in its sealed packaging until you are ready to install it.
Make sure that the o-ring is not pinched or rolled inside the cap.
The cap's lifetime is 1 year after the first reading has been taken.
Install by the date printed on the packaging.

The soft Cal Cup insert enables
calibration of the basic sensors
when the RDO sensor is installed.
It also contains an extra screw for the
adapter plate.

Adapter

Nose cone

RDO sensor

Restrictor

Lens wipe

Sensor cap

RDO sensor

Instruction sheet