Large Desiccant

Description
A clear acrylic tube with replaceable indicating desiccant, designed to protect a
vented Twist-Lock cable in high-humidity environments. Desiccates 20 times more
saturated air than the standard small desiccant. The desiccant gradually turns from
blue to pink as it absorbs moisture.

Material

Clear acrylic body, titanium or ABS connectors
(0051810 and 0053550, respectively), HDPE
membrane (high density polyethylene)

Environmental Rating

Weatherproof, not designed for submersion

Size

3.2 cm (1.25 in) O.D., 11.4 cm (4.5 in) long

Desiccant

approx. 22 g silica gel, type 4

Est. desiccant life (at 90% R.H.)

≥ 2 months to 50% saturation, depending on site
conditions

connections
1. Remove the protective dust cap from the bottom of the desiccant pack, if applicable.
2. Remove the small desiccant (if present) from the RuggedCable by grasping the

knurled (textured) section of the cable connector in one hand and the desiccant in the
other. Twist in opposite directions to unlock the desiccant from the cable.

3. Attach the large desiccant to the Twist-Lock connector on the cable.

Attaching cables and accessories is a simple matter of mating the Twist-Lock

connectors. This is easy after you've done it once or twice. For detailed steps and
photographs, see the product Quick Start Guide, or the Operator's Manual.

4. Remove the red dust cap to allow air to reach the cable's vent tube, if applicable.

guiDeLines & precautions
• To communicate with the instrument, remove the desiccant and attach a TROLL Com

to the cable.

• Desiccant will absorb moisture from the top down and for best results should be re-

placed before the entire volume has turned pink. A refill kit, Cat. No. 0029140, is avail-
able from In-Situ Inc. or your distributor.

• Protect new desiccant from moisture until ready to use.

Be sure to remove the red dust cap
from the top of the #0051810 desic-
cant before instrument deployment.