In-Situ RuggedCable for Level TROLL—Male connector end + stripped and tinned end User Manual

Short stripped & tinned cable, With male twist-lock connector, For level troll

background image

Information Sheet

Catalog No. 0053310

SHORT STRIPPED & TINNED
CABLE

with Male Twist-Lock Connector

Desiccant

Dust cap—Remove
before use

Strain relief

Twist-Lock
connector

A

B

C

Dust cap

Vent tube

and
secured to
cable with
strap

Outboard
desiccant
attached to
cable vent
tube

(over)

A

B

C

Level

TROLL

®

Accessories

Connects a Level TROLL's RuggedCable

to a controller or logger for communication

via:

• Analog (4-20 mA)

RS485 Modbus

• SDI-12

RS232 Modbus (with a customer-supplied converter)

Male Twist-Lock connector mates with female Twist-Lock connector on the RuggedCable

Stripped & tinned wires on opposite end for wiring to PLC or logger

Removable, refillable outboard desiccant protects cable vent tube from condensation in
high-humidity environments

SPECIFICATIONS
Materials

PVC, titanium, nylon, Viton

®

Dimensions

13.33 cm (5.25 in) long, 19 mm (0.75 in) O.D.

Environmental Rating

IP67 when connected, not designed for submersion

Cable

0.9 m (3 ft) vented TPU (Thermoplastic Polyurethane)

Conductors

6 conductors, 24 AWG, polypropylene insulation

Desiccant pack

Clear acrylic, nylon, HDPE membrane

Desiccant

approx. 22 g silica gel, type 4

Max. recommended

SDI-12

61 m (200 ft)

cable lengths

Modbus 1259 m (4000 ft)
4-20 mA 1259 m (4000 ft)

CONNECTIONS
Refer to wiring diagrams on back of sheet. Trim back and insulate unused wires. The
shield should be wired to a chassis ground or earth ground.

Power Connections. The Red wire provides power for Modbus and SDI-12 modes.
The Brown wire provides power for the 4-20 mA mode. If power is present on the Brown
wire and not on the Red wire, the device enters the 4-20 mA mode automatically and
stays in the 4-20 mode until power is removed from the Brown wire or is applied to the
Red wire. The Red wire has priority — if power is applied to both wires at the same time,
the device will operate in Modbus or SDI-12 modes but not in 4-20.

Communications. The device automatically switches between Modbus and SDI-12
modes depending on which of the two interfaces has activity. Modbus and SDI-12
cannot be used at the same time — whichever one is currently in use will block commu-
nication on the other.

To communicate with the device using Win-Situ

®

5 software through a serial COM port:

Disconnect the RuggedCable from the short stripped and tinned cable connector. Attach
a TROLL Com to the RuggedCable. Connect the TROLL Com's DB9-pin connector to a
serial port on a PC running Win-Situ 5.

TIP: The desiccant may be removed
from the vent tube, if needed, for
better access to the conductor wires.

Pull the vent tube extender off the cable vent
tube to remove. After connecting and trimming
wires, attach desiccant vent tube to cable vent
tube and secure desiccant to cable with hook-
and-loop strap, as shown here.