In-Situ RuggedCable for Aqua TROLL—Male connector end + stripped and tinned end User Manual

Short stripped & tinned cable, With male twist-lock connector, For aqua troll

background image

Information Sheet

Catalog No. 0053310

SHORT STRIPPED & TINNED
CABLE

with Male Twist-Lock Connector

Desiccant

Dust cap—Remove
before use

Strain relief

Twist-Lock
connector

A

B

C

Dust cap

Vent tube

and
secured to
cable with
strap

Outboard
desiccant
attached to
cable vent
tube

(over)

Aqua

TROLL

®

Accessories

for Aqua TROLL

®

200

Connects an Aqua TROLL's RuggedCable

or to a controller or logger for communica-

tion via:

• Analog (4-20 mA)

RS485 Modbus

• SDI-12

RS232 Modbus (with a customer-supplied converter)

Male Twist-Lock connector mates with female Twist-Lock connector on the RuggedCable

Stripped & tinned wires on opposite end for wiring to PLC or logger

Removable, refillable outboard desiccant protects cable vent tube from condensation in
high-humidity environments

SPECIFICATIONS
Materials

PVC, titanium, nylon, Viton

®

Dimensions

13.33 cm (5.25 in) long, 19 mm (0.75 in) O.D.

Environmental Rating

IP67 when connected, not designed for submersion

Cable

0.9 m (3 ft) vented TPU (Thermoplastic Polyurethane)

Conductors

6 conductors, 24 AWG, polypropylene insulation

Desiccant pack

Clear acrylic, nylon, HDPE membrane

Desiccant

approx. 22 g silica gel, type 4

Max. recommended

SDI-12

61 m (200 ft)

cable lengths

Modbus 1259 m (4000 ft)
4-20 mA 1259 m (4000 ft)

CONNECTIONS
Refer to wiring diagrams on back of sheet. Trim back and insulate unused wires. The
shield should be wired to a chassis ground or earth ground.

Power Connections. The Red wire provides power for all Aqua TROLL modes. The 4-
20 mA current loop output can be continuous in Modbus or SDI-12 mode as long as
Modbus device register 49507 is set to 1.

Communications. The device automatically switches between Modbus and SDI-12
modes depending on which of the two interfaces has activity. Modbus and SDI-12
cannot be used at the same time — whichever one is currently in use will block commu-
nication on the other.

To communicate with the device using Win-Situ

®

5 software through a serial COM port:

Disconnect the RuggedCable from the short stripped and tinned cable connector. Attach
a TROLL Com to the RuggedCable. Connect the TROLL Com's DB9-pin connector to a
serial port on a PC running Win-Situ 5.

A

B

C