In-Situ RuggedCable for Aqua TROLL—Female connector end + stripped and tinned end User Manual
Ugged, Able stripped & tinned, For aqua troll
Information Sheet
R
UGGED
C
ABLE
STRIPPED & TINNED
TM
*
Such as B&B Electronics Model 485SD9TB. See wiring on back of sheet
(over)
Aqua
TROLL
®
Accessories
Connects an Aqua TROLL directly to a controller or logger for communication via:
• Analog (4-20 mA)
• RS485 Modbus
• SDI-12
• RS232 Modbus (with a customer-supplied converter)
Twist-Lock connector mates with probe body. Stripped & tinned wires on opposite end
for wiring to PLC or logger. Vented cable includes removable outboard desiccant to
protect cable vent tube from condensation in high-humidity environments.
CABLE SPECIFICATIONS
Jacket options
TPU (Thermoplastic Polyurethane)
Halogen-Free TPU (LSZH-rated, low smoke zero halide)
FEP (fluorinated ethylene propylene, generic equivalent of
DuPont Teflon
®
)
Vent options
Vented (includes outboard desiccant), non-vented
Conductors
6 conductors, 24 AWG, polypropylene insulation
Diameter
Cable, 6.7 mm (0.265 in); Connector, 18.5 mm (0.73 in)
Minimum bend radius
2X cable diameter (13.5 mm, 0.54 in)
Break strength
127 kg (280 lb)
Max. recommended
SDI-12
61 m (200 ft)
cable lengths
Modbus
1259 m (4000 ft)
4-20 mA
1259 m (4000 ft)
Desiccant pack
Clear acrylic, nylon, HDPE membrane
Desiccant
approx. 22 g silica gel, type 4
CONNECTIONS
Refer to wiring diagrams on back of sheet. Trim back and insulate unused wires. The
shield should be wired to a chassis ground or earth ground.
Power Connections. The Red wire provides power for all Aqua TROLL modes. The 4-
20 mA current loop output can be continuous in Modbus or SDI-12 mode as long as
Modbus device register 49507 is set to 1.
Communications. The device automatically switches between Modbus and SDI-12
modes depending on which of the two interfaces has activity. Modbus and SDI-12
cannot be used at the same time — whichever one is currently in use will block commu-
nication on the other.
To communicate with the Aqua TROLL using Win-Situ
®
5 software through a serial COM
port: Cable can be wired to a port-powered RS485-RS232 converter* for temporary
connection to a serial port on a PC.
Signal
Color
Gnd/Return
BLACK
Ext Power
RED
4-20 mA
BROWN
RS485(–)
GREEN
RS485(+)
BLUE
SDI-12
WHITE
RuggedCable (TPU)
Signal
Color
Gnd/Return
BLACK
Ext Power
RED
4-20 mA
BROWN
RS485(–)
GREEN
RS485(+)
ORANGE
SDI-12
YELLOW
FEP Cable
and
secured to
cable with
strap
Outboard
desiccant
attached to
cable vent
tube
Dust cap
Remove
before use
TIP: The desiccant may be removed
from the vent tube, if needed, for
better access to the conductor wires.
Pull the vent tube extender off the cable vent
tube to remove. After connecting and trimming
wires, attach desiccant vent tube to cable vent
tube and secure desiccant to cable with hook-
and-loop strap, as shown in the photo above.
for Aqua TROLL
®
200