About 3V (for 200Ω range);

1. 30 ranges for function selection;

2. Over-range display: "1";

3. Display count: 19999;

4. Fast-acting fuse protection for current ranges: 200mA and 10A;

5. Full-range overload protection;

6. Auto power off;

7. Temperature range:

Operating: 0℃~ 40℃(32℉ ~ 104℉);

Storage: -10℃~ 50℃(14℉ ~ 122℉);

8. Altitude: ≤ 2,000m;

9. Low Battery Indication: “    ” displays;

10. Dimensions: 186Ч91Ч39mm;

11. Weight: About 300g (not including test leads).

Ⅰ.

Introduction

Ⅲ.

Product Features

Ⅳ.

Technical Specifications

The GDM-452 is a 20000-count handheld digital multimeter

with remarkably stable and reliable operation. It is mainly

supported with a dual integral A/D converter based on CMOS technology

and also offers overload protection for all ranges. This model can

measure DC&AC voltage, DC&AC current, resistance, capacitance,

diode, transistor, frequency and continuity.

Ⅱ.

Safety Information

The GDM-452 is designed and manufactured in compliance with:

IEC61010-1, CATⅠ1000V, CATⅡ600V and CATⅢ300V overvoltage

standards.

Warnings

Use the meter as specified in the manual , otherwise the protection

offered by the multimeter may be impaired.

● Do not use the multimeter with back cover opened, it may cause

electric shock.

● Set to a proper range for your measurement.

● Check the test leads for any damaged insulation or broken wires.

● Insert red and black test leads into proper input terminals and ensure

good contact between them.

● Do not input signal beyond the rated values of the multimeter,

otherwise it may cause electric shock or damage to the multimeter.

● Prohibit switching the range during the voltage or current

measurement, for it may cause damage to the multimeter.

● Please use replacement parts with the same model or similar electrical

specifications.

● To avoid electric shock, Do not apply any voltage above 1000V

between COM and grounding.

● Please take caution when working voltage goes above 60VDC or

30VAC RMS.

● To ensure the accuracy, please replace the batteries as soon as      

appears.

● Shut off the power timely after the measurements complete. Take out

the batteries if not used for a long time.

● Do not use the multimeter in places exposed to high temperature, high

moisture. The performance of the multimeter may be compromised if

moisture-affected.

● To prevent damage to the multimeter or personal injury, do not alter

internal wiring randomly.

● Clean the multimeter casing with slightly damped soft cloth and mild

agent. Do not use any abrasives and corrosives.

International Electrical Symbols

Low Battery Indication

Double Insulated

Conforms to European Union Standards

b digits

DC Voltage

AC Current

Resistance

Capacitance

Frequency

Diode and Continuity

AC Voltage

DC Current

AC RMS

AC RMS

True RMS v

RMS

Note: please short-circuit the test leads when using 200 Ω range to

measure, and subtract this shorted value from all subsequent measured

values so as to obtain accurate reading.

Display approximate value

of forward voltage drop for

diode

If resistance≤30Ω, the

buzzer sounds; Display

approximate kΩ value.

GDM-452

Effective average value