FREQUENCY COUNTER

USER MANUAL

8

z Front Panel

Fig. 1 Front panel

FREQUENCY COUNTER

USER MANUAL

9

5. APPLICATION

(1) Signal Measurements

Frequency range between 0.01Hz and 120MHz, press FREQ A switch,

then connect input signal to the channel A input BNC.

Frequency range between 50MHz and 1.3(2.7) GHz, press FREQ B

switch, then connect input signal to the channel B input BNC.

Press “PRID” switch to select period mode of operation for signal on

input A.

(2) Gate Time Settings

The instruments features continuously adjustable gate time selection

from 10ms to 10s or one period of input, depending on whichever is

longer. The GATE TIME adjustment affects the sampling rate and the

resolution of the reading.

Turn the knob counterclockwise for updating faster reading, or turn it

clockwise for more resolution (digital display).

Pulling out the GATE TIME knob to freeze the latest displayed reading

for being written down. Then return the counter to normal operation by

pressing back the knob.

When the instrument is starting measurement, the LED indicator above

the GATE TIME knob will light. In normal operation, it will blink at a

rate setting by GATE TIME knob.

(3) Trigger Level Adjustment

Adjust the trigger voltage of channel A input signal by pulling and

turning the TRIG LEVEL knob for triggering. The trigger voltage is

variable over ±2.5V×ATT setting.

If push the knob, it can be functioned as auto-triggering.(this knob is

only available for channel A.)