Detcon P-1000 User Manual
Page 11
P-1000
Portable Model 1000 H2S Analyzer
Rev. 1.0
Page 7 of 24
7. Click the Launch icon
on the toolbar. This displays the logger’s Launch window. Launch
windows vary for each type of logger, but most should look similar to Figure 6.
8. Review the default Launch settings. Enter an appropriate name in the Description field, (Test1 in this
example) and select Channel 1in the Channels to Log select box.
Figure 6 HOBO Launch Window
9. Select Scaling to open the Linear Scaling Window (Figure 7). The scale channel check box should be
checked. Give the channel an appropriate name, (H2S in this example) and set the scaling units to ppm
with 4mA set for 0.0 and 20mA set for full scale. I.e. for 0-20ppm full scale would be 20ppm, or
20mA=20ppm. Select OK to return to the Launch Logger Screen.
Figure 7 Set Scaling
10. Almost any logging Interval can be set from 1 second to over 18 hours, it is suggested to set a 1 to 5
minute Logging Interval. Choose the Now launch option to immediately start-logging data, other options
may be selected and are discussed in detail in the HOBOware™ User’s Guide. Click Launch to begin