2.2 Program Mode

While in Normal Operation with Concentration displaying, pressing the “ENT” pushbutton and holding it for
about three seconds allows the Program Mode to be accessed. While in the Program Mode, the left
pushbuttons have the meaning “×”and “Ø”which allow moving back and forth or up and down through the
menu and for changing values up and down. If an alarm triggers while in the Program Mode, the 10C
Controller will automatically exit the Program Mode and return to Normal Operation whereupon the up and
down pushbuttons become the alarm reset/silence pushbuttons. The Program Mode allows setting the many
variables in the 10C Controller:

Set Gas Type

Set

Range

Set

Units

Set Alarm 1:

Set-Point

Direction

Latching

Energized

Silenceable

Set Alarm 2:

Set-Point

Direction

Latching

Energized

Silenceable

Set

Fault:

Set-Point Threshold in milliAmps and always descending

Latching

Energized

Silenceable

Set Modbus™ Address

Set Sensor-Cal Mode

Test

Trim Input (Technician Access only)

Trim Output (Technician Access only)

Restore Default Settings (Technician Access only)