Operator interface, Normal operation, 0 operator interface – Detcon 10C User Manual
Page 11: 1 normal operation

Model 10C
2.0 Operator Interface
The 10C Controller is configured by use of the front panel display and pushbuttons. The 10C Controller User
Interface Menu System allows the user to set various parameters associated with the sensor: Gas type, Range,
Units, Alarm settings, and Modbus ID.) Navigation through the menu system is accomplished by use of the
10C Controller’s Front Panel Pushbuttons: the Enter key “ENT”, the Escape key “ESC”, the Up key “×” (or
“RESET”), and the Down key “Ø” (or “SLNC”). The menu system includes the main Programming menu and
submenus as indicated in this section.
2.1 Normal Operation
Normal Operation of the display is to constantly display the Concentration Reading. Upon power-up
initialization of the 10C Controller, the text “WARM UP” is displayed instead of concentration for about sixty
seconds. While in the Normal Operation mode, the Pushbuttons on the left have the meaning “Alarm Reset”
and “Alarm Silence” and are used to Reset and to Acknowledge alarms.
In Normal Operation the Gas Type and Units can be prompted for momentary display by pressing the “ESC”
pushbutton. The Program Configuration Status can be viewed by pressing the “ENT” pushbutton:
View Program Status:
(Lamp
Test)
Gas
Type
Range
Units
Alarm 1 Set-Point
Alarm 1 Direction
Alarm 1 Latching
Alarm 1 Energized
Alarm 1 Silenceable
Alarm 2 Set-Point
Alarm 2 Direction
Alarm 2 Latching
Alarm 2 Energized
Alarm 2 Silenceable
Fault
Set-Point
Fault
Direction
Fault
Latching
Fault
Energized
Fault
Silenceable
Modbus™
Address
Firmware
Version
Firmware
Checksum
Firmware last programmed address (Top of code address in memory)
10C Control Module Instruction Manual Rev. 0.B
Page 7 of 25