Detcon IR-542 User Manual

Page 8

background image

3.4.3 Program Mode

The program mode provides a program status menu, allows for the adjustment of alarm set point levels, and the

selection of the calibration gas level setting..

3.4.3.1 Program Status

The program status scrolls through a menu that displays:

* The gas type, range of detection and software version number. The menu item appears as:

“CO2 RANGE V6.0”

* The calibration gas level setting. The menu item appears as:

“CalLevel @ #.##%”

* Sensor temperature in °C appears as:

“TEMPERATURE ## °C”

* The estimated remaining sensor life. The menu item appears as:

“OPTICS AT 100%”

3.4.3.2 Calibration Level Adjustment

The Calibration level is adjustable from 10% to 90% of full-scale range. The menu item appears as:

“CalLevel @ #.##%”

3.5 I

NSTALLATION

Optimum performance of ambient air/gas sensor devices is directly relative to proper location and installation practice.

3.5.1 Field Wiring Table (4-20 mA output)

Detcon Model IR-540/IR-541/IR-542 CO2 sensor assemblies require three conductor connection between power

supplies and host electronic controllers. Wiring designators are + (DC), (DC) , and mA (sensor signal).

Maximum single conductor resistance between sensor and controller is 10 ohms. Maximum wire size for termina-

tion in the sensor assembly terminal board is 14 gauge.

AWG

Wire_Dia.

Meters

Feet

20

0.812mm

1120

3350

18

1.024mm

1750

5250

16

1.291mm

2800

8400

14

1.628mm

4480

13,400

Note 1: This wiring table is based on stranded tinned copper wire and is designed to serve as a reference only.

Note 2:

Shielded cable may be required in installations where cable trays or conduit runs include high voltage

lines or other sources of induced interference.

Note 3:

The supply of power must be from an isolating source with over-current protection as follows:

AWG

Over-current Protection

AWG

Over-current Protection

22

3A

16

10A

20

5A

14

20A

18

7A

12

25A

The RS-485 (if applicable) requires 24 gauge, two conductor, shielded, twisted pair cable between sensor and host

PC. Use Belden part number 9841. Two sets of terminals are located on the connector board to facilitate serial loop

wiring from sensor to sensor. Wiring designators are A & B (IN) and A & B (OUT).

3.5.2 Sensor Location

Selection of sensor location is critical to the overall safe performance of the product. Five factors play an important

role in selection of sensor locations:

(1) Density of the gas to be detected

(2) Most probable leak or generation sources within the industrial process

(3) Ventilation or prevailing wind conditions

(4) Personnel exposure

(5) Accessibility for routine maintenance

Detcon Model IR-540/IR-542 Carbon Dioxide Sensor PG.8