beautypg.com

13 s – Detcon IR-542 User Manual

Page 17

background image

Bad 4-20 mA output

1. Check that wiring is connected to correct terminal outputs.

2. Swap with a known-good transmitter to determine if transmitter is faulty

Unstable output/Sudden Spiking/Nuisance Alarms

1. Check condulet for accumulated water.

2. Check transmitter and Terminal PCB for abnormal corrosion.

3. Determine if problem correlates with condensation cycles.

4. Add/change Detcon condensation prevention packet.

5. Check for unstable power supply.

6. Check for inadequate grounding.

7. If correlates with radio communications then use Detcon RFI filter accessory.

8. Contact Detcon for assistance in optimizing shielding, grounding, and RFI protection.

Excessive Span Drift or Slow Response

1. Verify correct cal gas flow rate and proper use of the cal gas adapter.

2. Check validity of cal gas via the expiration date and use pull tube if necessary.

3. Check for obstructions through s/s sinter element (including being wet).

4. Replace plug-in sensor if Optics Life is <25%.

Drifting Zero

It may be the correct reading if there are real gas leaks or the sensor was zero calibrated when actual gas was present and subse-

quently cleared.

1. Recalibrate zero using target gas “free” sample.

2. Replace plug-in sensor if Optics Life is <25%.

3.13 S

PARE

P

ARTS

L

IST

613-010000-000

Sensor rain shield

613-120000-000

Sensor splash guard

943-000006-132

Threaded Calibration Adapter

390-000088-000

IR sensor housing assembly (plug-in detector PN370-365878-111 or 370-287724-232,

not included)

370-365878-111

Field replaceable optical plug in sensor (for IR-540 .3 to 5% range)

370-365878-222

Field replaceable optical plug in sensor (for IR-541 .3 to 5% range)

370-287724-232

Field replaceable optical plug in sensor (for 10 to 100% range)

925-405400-xxx

IR-540 Plug-in control circuit (where xxx is 0X3 = 0.30%, 0X5 = 0.50%, 001 = 1%,

1X2 = 1.2%, 003 = 3%, and 005 = 5%)

925-415400-xxx

IR-541 Plug-in control circuit (where xxx is 0X3 = 0.30%, 0X5 = 0.50%, 001 = 1%,

1X2 = 1.2%, 003 = 3%, and 005 = 5%)

925-425400-xxx

IR542 Plug-in control circuit (where xxx is 010 = 10%, 025 = 25%, 050 = 50%, 100 = 100%)

500-002226-001

IR Connector board

327-000000-000

Programming Magnet

897-850800-000

3 port enclosure less cover

897-850700-000

Enclosure glass lens cover

960-202200-000

Condensation prevention packet (replace annually).

Detcon Model IR-540/IR-542 Carbon Dioxide Sensor PG.17

This manual is related to the following products: