16 s, Oftware, Hart – Detcon IR-542 User Manual
Page 19: Normal operation

3.16 S
OFTWARE
F
LOW
C
HART
Detcon Model IR-540/IR-542 Carbon Dioxide Sensor PG.19
NORMAL
OPERATION
CALIBRATION
PGM1 (3)
PGM2 (3)
1-ZERO 2-SPAN
LEGEND
PGM1 - program switch location #1
PGM2 - program switch location #2
(M) - momentary pass of magnet
(3) - 3 second hold of magnet
(30) - 30 second hold of magnet
INC - increase
DEC - decrease
# - numeric value
AUTO ZERO
AUTO SPAN
LEL 0-100 V#.#
PGM1 (3)
PGM2 (M)
VIEW PROG STATUS
PGM2 (3)
CAL LEVEL @ ##%
PGM1 (3)
PGM1 (M)
PGM2 (30) PGM2 (M)
OPTICS AT ##%
TEMPERATURE ##∞C
PGM1 (M) PGM2 (M)
PGM1 (3)
CAL LEVEL @ ##%
INC
DEC
PGM1 (3)
PGM2 (M)
SET CAL LEVEL
PGM2 (3)