Figure 15 fp configuration setup – Detcon TP-524D-HRT User Manual
Page 35
Model TP-524D-HRT
TP-524D-HRT Instruction Manual
Rev. 0.3
Page 31 of 48
Figure 15 FP Configuration Setup
The Range Set up will display the Model Type of the sensor, followed by the PV LRV (Primary Variable
Lower Range Value) and the PV URV (Primary Variable Upper Range Value), and the Conc Units
(Concentration Units), the display may also show the Sensor Range. The Sensor Setup portion of the screen
will display sensor specific parameters:
Heater Power
Heater Voltage
Sensor Resistance
Heater Current
Sensor Life
NOTE: The values above are read when the HRT Bridge boots up and are not updated
in real-time.
4.4.3.2 Calibration
The Calibration screen displays the days since the last calibration, and the auto span level. This screen also
allows the user to calibrate the sensor by performing an Auto Span Calibration. Calibration of the sensor using
this feature also notifies the Host that the sensor is in calibration mode.
TP