Detcon TP-524D-HRT User Manual
Page 21
Model TP-524D-HRT
TP-524D-HRT Instruction Manual
Rev. 0.3
Page 17 of 48
View Sensor Status displays all current configuration and operational parameters including: sensor type,
software version number, detection range, AutoSpan level, days since last AutoSpan, estimated remaining
sensor life, heater power, heater voltage, raw resistance, mA output, input voltage, and sensor ambient
temperature.
From the View Sensor Status text scroll, hold the magnet over PGM1 or PGM2 until the “
” prompt appears
and continue to hold the magnet in place for an additional 3-4 seconds (until the display starts to scroll “Status
Is”). The display will scroll the complete list of sensor status parameters sequentially:
Sensor Model Type
The menu item appears as: “TP-524D”
Current Software Version
The menu item appears as: “V X.XXZ”
Range of Detection.
The menu item appears as: “Range XXXppm”
AutoSpan Level.
The menu item appears as: “Auto Span Level XXppm”
Days Since Last AutoSpan.
The menu items appears as: “Last Cal XX days”
Remaining Sensor Life.
The menu item appears as: “Sensor Life 100%”
Sensor Heater Power.
The menu item appears as: “Heater XXXmW”
Sensor Heater Voltage
The menu item appears as: “Heater X.XXVDC
Raw Sensor Resistance
The menu item appears as: “Resistance XXXXX”
mA Output
The menu item appears as: “mA Output XX.XX mA”
Input Voltage Supply
The menu item appears as: “Voltage XX.X VDC”
Operating Temperature
The menu item appears as: “Temp XX C”