Setup menu – Delmhorst Instrument Moisture Check User Manual
Page 20
Copyright
2005, Delmhorst Instrument Co.
Page 20 of 29
c. Sync: This message occurs when data framing is lost during data transmission. This
would only happen if the wireless link becomes unreliable. Ensure the meter and
base unit are aligned and restart the session.
d. Data E: This message occurs when the content of a correctly framed data packet is
invalid. Again, this is an unreliable wireless link problem.
NOTE: Individual records cannot be downloaded to the PC. The entire storage content is
delivered to the PC during a link session
SETUP MENU:
The Setup group of features allows access to the configurable parameters of the meter. They are:
Date and Time, Power-off time, Backlight, and Conversions (temperature mode and electrode
type). All can only be accessed in the Setup Menu. To get to the Setup Menu, use the
с т
keys to highlight option #5:
SETUP.
SETUP MENU ENTRY DESCRIPTIONS:
o
Return to Main Menu: This menu option returns the meter to Main Menu. The meter
also returns to Main Menu if no user input is detected for 20 Seconds.
o
Set Date and Time: This menu option allows user access to the date and time set
screen:
The Date and Time set screen is populated with the time and date values present in the
meter at access time. UP and DOWN key action switches the row selection among available
parameters. LEFT and RIGHT key action changes the selected parameter’s value (for
“DONE” and “CANCEL” – parameters without assigned values, the LEFT and RIGHT keys
perform no function). All parameter values are incremented using the RIGHT key and
decremented using the LEFT key. The DAY parameter moves forward through weekdays
using the RIGHT key and backward using the LEFT key. The SELECT (center) key is only
active when menu selection is applied to “HOURS” (switch between standard or military
time), “DONE” or “CANCEL”. Selecting “DONE” exits this screen back to the SETUP
MENU and updates the clock with the newly chosen values. Selecting “CANCEL” exits this
screen back to the SETUP MENU without updating the clock.