31

Technical Data

Technical Data

Conforms to: ISO 8502-4, BS7079-B, ASTM D3276,

US Navy NSI 009-32, IMO PSPC, SSPC-PA 7 and others.

Range

Accuracy

Resolution

Surface
Temperature

-40° to 80° C

80° to 190° C

-40° to 175° F

175° to 375° F

±0.5° C
±1.5° C
±1° F
±3° F

0.1° C
0.1° C

0.1° F
0.1° F

Air
Temperature

-40° to 80° C

-40° to 175° F

±0.5° C

±1° F

0.1° C

0.1° F

Humidity

0 to 100%

±3 %

0.1 %

Operating
Range

-40° C to +80° C

-40° F to +175° F

the Gage was powered-on and the Gage was not able to self
power-down. If this message appears it will be followed by a full
reset. All datasets in memory will be erased and Gage settings will
be reset to "out-of-the-box" settings.

Gage will not print to the Bluetooth wireless printer:

- Ensure the Gage has been paired for use with the printer and
the Gage menu option Connect > Bluetooth > Stream box is
ticked.
- Check printer batteries. Switch it off and then on again.