Technical Data
Technical Data
Conforms to: ISO 8502-4, BS7079-B, ASTM D3276,
US Navy NSI 009-32, IMO PSPC, SSPC-PA 7 and others.
Range
Accuracy
Resolution
Surface
Temperature
-40° to 80° C
80° to 190° C
-40° to 175° F
175° to 375° F
±0.5° C
±1.5° C
±1° F
±3° F
0.1° C
0.1° C
0.1° F
0.1° F
Air
Temperature
-40° to 80° C
-40° to 175° F
±0.5° C
±1° F
0.1° C
0.1° F
Humidity
0 to 100%
±3 %
0.1 %
Operating
Range
-40° C to +80° C
-40° F to +175° F
the Gage was powered-on and the Gage was not able to self
power-down. If this message appears it will be followed by a full
reset. All datasets in memory will be erased and Gage settings will
be reset to "out-of-the-box" settings.
Gage will not print to the Bluetooth wireless printer:
- Ensure the Gage has been paired for use with the printer and
the Gage menu option Connect > Bluetooth > Stream box is
ticked.
- Check printer batteries. Switch it off and then on again.