beautypg.com

COSA Xentaur Optical Flow Meter User Manual

Page 8

background image

8

Making Light Work

~ Photon Control Inc

.

3.3 Procedure – Calculating Insertion Depth – 4” Pipe:

Before progressing with the installation of the probe it is necessary to calculate and mark the
correct insertion depth. The actual measuring point for the Focus TM Probe is 1.65” (42mm)
from the end of the probe. For a 4” pipe diameter this measuring point should be positioned at
the center point for accurate readings. This is calculated using the following method. Refer to
Figure 5 for a diagram of the measurements to be taken.

1. Measure and record distances A, B, C and D.

2. Calculate and record distance E = A+B+C+D. (Keep this record for future reference as it will be

needed for reinstallation should the probe be removed.)

3. Mark the probe at distance E from the probe end as shown.

M

easure and record the following:

A = End to Measuring Point

= 1.65” +

B = 0.5 x Pipe ID

= _____ +

C = Pipe Wall Thickness

= _____ +

D = Outer Wall to Packing Gland

= _____ =


E = Insertion Depth = A+B+C+D

= _____

Figure 6: Insertion Depth Calculations for 4" and Smaller Pipe Diameters

NOTE:

Pipe internal diameter and wall
thickness may be found in

Appendix A.

NOTE:

Insertion depth tolerance:

±0.5%

of nominal pipe size.