0 focus, Probe maintenance – COSA Xentaur Optical Flow Meter User Manual
Page 13
Photon Control Inc
.
~
Making Light Work
13
5.0 Focus
TM
Probe Maintenance
1. Check that alignment scribe marks are present on the flow meter as shown in Figure 13.
Figure 13: Alignment Scribe Marks on Insertion Probe Components
2. Disassemble the 3 main components of the flow meter as shown in Figure 14 by unscrewing
counterclockwise.
3. Inspect o-rings for wear and cracks and replace as necessary.
4. Inspect all thread surfaces for burrs and dirt. Deburr and clean thoroughly before reassembly.
Figure 14: Disassemble 3 Main Components to Access Optics for Cleaning
Inspect o-rings and replace if
cracked, split or worn.
Alignment Scribe Marks.
Deburr and clean threads
before reassembly
NOTE:
Ensure that scribe marks are
aligned and that L2 is toward the
tip on reassembly.