Making Light Work

~ Photon Control Inc

.

3.0 Focus

TM

Probe Installation

The instructions in this section cover the installation of the Focus

TM

Probe and the safety hardware used

to retain the probe. Figure 4 through

Figure 6: Insertion Depth Calculations for 4" and Smaller Pipe

Diameters

provide visual aids for the installation procedures.

At no time during the installation should any person position any part of their body in line with the probe.
A pressurized pipe could cause the probe to accelerate outwards causing bodily harm.

Small leakages may occur during installation. Take appropriate safety precautions.

Install only under minimal operating pressure or with the line shut off to prevent large forces acting on the
probe.






3.1 Procedure – Installing the Safety Chain Clip:

1. Undo the Quick Link connector and remove the Safety Chain from the Safety Chain Clip.

2. Slide the Safety Chain Clip on to the probe so the tab points away from the Fiber Connector at the

end of the probe.

3. Slide a Clamp-On Collar on to the probe and position it so it secures the Safety Chain Clip against

the Fiber Connector.

4. Tighten the screw on the Clamp-On Collar with a 3/16” Allen wrench.

Figure 4: Install Safety Chain Clip and Clamp-On Collar as Shown

NOTE:

Maximum Force Acting on Probe =

Line Pressure (psig) x 0.44

Examples:

30psig x 0.44 = 13lbs

100psig x 0.44 = 44lbs

NOTE:

Process operating pressure and temperature
restrictions:


Max Pressure: 100psig

Temp. Range: -40ºF to +302ºF (-40ºC to +150ºC)

SAFETY CHAIN CLIP

CLAMP-ON COLLAR

FIBER CONNECTOR

SAFETY CHAIN