2.0 Focus

TM

Probe Pipe Fitting Installation – Hot Tap

1. Choose the location in the pipe that you want to install the Focus

TM

Probe.

2. Install a Threadolet with a 1” NPT thread.

3. Apply an appropriate pipe sealant and thread the 1” to 1” Nipple into the Threadolet and tighten.

4. Apply an appropriate pipe sealant and, holding the Ball Valve so the mounting holes and lever

point away from the pipe, thread the 1” Ball Valve on to the Nipple and tighten.







Figure 2: Hot Tap Fitting Installation

5. Hot Tap Drill a 7/8” hole through the pipe wall.

6. Turn the Ball Valve to the “Off” position (handle perpendicular to valve body) when removing the

Hot Tap Drill from the pipe fittings.

7. Slide the Hot Tap Tool into the Packing Gland with the dowel pin protruding on the threaded end.


PIPE

1” TO 1” NIPPLE

1” NPT ISOLATION VALVE

1” Threadolet

NOTE:

1) Apply sealant to NPT

threads before installing.

NOTE:

Internal threads on 1” to 1” nipple

towards pipe

INTERNAL THREADS