7. What do I do if…
… no weight display comes on?
– Is the scale switched on?
– Check the batteries.
– Is the plug contact for connecting the display to the scale correctly connected?
…
0.00
does not appear before weighing?
– Switch off the scale with the Start key and start the scale again – there must not be
any load on the scale – and only its feet should be in contact with the floor.
… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
… the display shows
?
– Battery voltage is running low. Change the batteries in the next few days.
…
bAtt
appears in the display?
– Batteries are empty. Put in new batteries.
…
St0p
appears in the display?
– Maximum load has been exceeded.
… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
and wait until
0.00
is displayed, then weigh again.
… the display
appears?
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
… the display
Er
and a number appear (e.g.
Er:0:12
)?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the batteries. If
this measure is equally unsuccessful, inform the Service department.