Store weight value (HOLD)
You can also continue displaying the weight determined once the load has been removed.
This makes it possible to look after a patient first, before having to note the weight:
– Press the
HOLD/TARE key whilst the
load is on the scale.
“HOLD” appears in the digital display. The
value remains saved in the display until the
hold function is automatically switched off.
– The HOLD function can be switched off
by pressing the
HOLD key again.
0.00
appears in the display if the scale has
no load on it, otherwise the current weight.
The “HOLD” display vanishes.
Note:
The HOLD function can also be used if the TARE function has been executed beforehand.
Switch Auto-Hold function on/off
The HOLD function is activated automatically at weights of over 2.5 kg. The display flash-
es until a stable weight has been measured. The display is then frozen until the next
weighing operation takes place. HOLD appears in the display.
You can also switch this function off:
– Press and hold the
FUNC/SET key for
approx. 2 seconds.
The Settings menu is now selected.
– Press one of the arrow keys briefly until
k0Ld
appears in the display.
– Confirm your selection by pressing-
FUNC/SET.
The current setting appears in the dis-
play:
0n
= Auto-Hold switched on,
0ff
= Auto-Hold switched off.
– Press one of the arrow keys briefly until
the desired setting appears in the dis-
play.
– Confirm your selection by pressing
FUNC/SET to save the setting dis-
played.