beautypg.com

Store weight value (hold), Switch auto-hold function on/off – Seca 703 User Manual

Page 22

background image

22

Store weight value (HOLD)

You can also continue displaying the weight determined once the load has been removed.

This makes it possible to look after a patient first, before having to note the weight:
– Press the

HOLD/TARE key whilst the

load is on the scale.

“HOLD” appears in the digital display. The

value remains saved in the display until the

hold function is automatically switched off.
– The HOLD function can be switched off

by pressing the

HOLD key again.

0.00

appears in the display if the scale has

no load on it, otherwise the current weight.

The “HOLD” display vanishes.
Note:
The HOLD function can also be used if the TARE function has been executed beforehand.

Switch Auto-Hold function on/off

The HOLD function is activated automatically at weights of over 2.5 kg. The display flash-
es until a stable weight has been measured. The display is then frozen until the next

weighing operation takes place. HOLD appears in the display.
You can also switch this function off:
– Press and hold the

FUNC/SET key for

approx. 2 seconds.
The Settings menu is now selected.

– Press one of the arrow keys briefly until

k0Ld

appears in the display.

– Confirm your selection by pressing-

FUNC/SET.
The current setting appears in the dis-
play:

0n

= Auto-Hold switched on,

0ff

= Auto-Hold switched off.

– Press one of the arrow keys briefly until

the desired setting appears in the dis-
play.

– Confirm your selection by pressing

FUNC/SET to save the setting dis-
played.