Maintenance 21

6 Maintenance

6.1 Cavity Life

The life expectancy of the cavity depends on the time and temperature of operation. The higher
the temperature, the shorter the life span. At temperatures near the upper end of the range,

the life expectancy is shortened by approximately 30%.

6.2 Safety Interlocks

Automatic shut-down will occur if the region near the cavity housing exceeds ~115°C. This
thermal fuse must be replaced when over-temperature causes a shut down. The cooling fan will
operate when power is on.

6.3 Periodic Inspections

Check periodically and upon every start up the following items:
1.

Unusual noises and smells when applying power.

6.4 Fuse Locations

The M310 has one or two fuses. Refer to the table below.
Fuse Nµmber Amps/volts

Location

Description

F1 (120V)

8A / 250 VAC Fast External-Rear Panel Integral with IEC power connector

F2

(120V)

NONE

NONE

NONE

F1 (230V)

4A / 250 VAC Fast External-Rear Panel Integral with IEC power connector

F2 (230V)

4A / 250 VAC Fast

6.5 Cleaning

Clean the outside of the cabinet with mild soapy solution and water only. Do not attempt to
clean the emitter plate.

6.6 Heater Element Replacement

The M310 heating element is not field replaceable since the cavity and elements are an integral

assembly. Contact LumaSense for service if the heating element fails. (Check continuity of the
thermal fuse first.)