Communications Port 18

RS485 wiring

The blackbody was shipped with the controller set at the following com port parameters:

Baud: 1200

Data Bits: 7

Stop bits: 1

Parity: odd

To alter these parameters, see Section 4.5.

5.3 Communication Command Set

The complete command set is extensive. For the purpose of brevity, only a few important ones
are shown here. For more information, the controller instruction manual may be requested from
LumaSense.

Command line to read current blackbody temperature:
Controller ID Parameter #

checksum

$0101R05C1

end of message

read command

zone #

start of message

Controller response:
Controller ID Read command

data

%0101R0502.000J7

checksum

zone # parameter #

start of message

Command line to read current blackbody setpoint:

$0101R09C5