M335 Manual

Introduction 9

2 Introduction

2.1 Description

The M335 blackbody calibration source is a general purpose high temperature source which
provides a very rapid heat-up time of only 30 minutes to reach 1200 C. A self-tuning digital PID
controller with adjustable set point holds the temperature to within 1 C.
The Model M335 blackbody calibration source design achieves significant weight and size
reduction over the M330 series with only minor sacrifices in specification and performance,
making it an ideal unit where portability is required or space is limited. The cavity combined
with radiance calibration yields an emissivity of 1.0 over the temperature range of 300 to

1500 °C with an aperture of 16mm. An integrally mounted PID controller completes this versatile
calibrator.

2.1.1 Accuracy

Typical accuracy of the after radiometric calibration is ±0.4% ±1 °C of reading. Please refer to
the NIST traceable calibration report for the measurement uncertainties and controller setpoint
corrections (systematic corrections).
The M335 Blackbody Calibration Furnace has been thoroughly tested at factory prior to
shipment.
Every 300 to 400 hours of operation check the accuracy of temperature calibration at several
points throughout the range. This test is very important because the control thermocouple will
gradually lower its output causing the temperature to increase. A running log of the calibration
should be maintained by your standard department.

2.2 Technical Data

Temperature Range:

300 to 1500 °C or 570 to 2730 °F

Accuracy:

±0.4% of reading ±1 C

Temperature Resolution:

1 C

Stability:

1.0 C per 8 hour period

Aperture Diameter:

16.5 mm (0.65")

Source Non-Uniformity:

Non-Uniformity: ±1 C within center 1/3 of ID.

Heated Cavity Shape:

Closed end tube

Effective Emissivity

1.0 at 0.65-1.8 µm

Sensor:

Precision platinum thermocouple Type B

Method of Control:

Digital Self-Tuning PID Controller

Warm-up Time:

30 minutes from ambient to 1200 °C.

Operating Ambient Temp.:

0 C to 44 C (32 F to 110 F)

Cooling:

Fan cooled, air inlet on rear panel

Power Requirements:

115VAC±10% 50/60 Hz. 2.0kw max. (230 VAC optional)

Dimensions:

290 mm H x 495 mm W x 550 mm D

Maximum Weight:

28 kg (62 lbs)