1 other controller modes, 3 shutting down, 4 changing temperature units – LumaSense Technologies M335 User Manual
Page 12
M335 Manual
Operation 12
Eurotherm controller
3.2.1 Other Controller Modes
If, for some reason, the power must be turned off while monitoring temperature, you may
manually set the power to none by pressing the AUTO, then raise or lower the power level in
the same manner used to change the setpoint. Set to 0% to apply NO power to the blackbody.
Warning: NEVER SET THE POWER TO ANY LEVEL EXCEPT 0% MANUALLY. If left
unattended, the blackbody will overheat and be damaged.
To return to normal operation, press AUTO/MAN again.
3.3 Shutting Down
Turn off power using the power ON/OFF switch.
Warning: DO NOT UNPLUG THE M335! The blackbody has a temperature
controlled cooling fan that remains active until the enclosure temperature is
below about 140 °F. If the power is interrupted by external means, this cooling will
not occur and the devices in the enclosure may overheat. If power fails during high
temperature use, remove the top, rear cover plate to allow better convection cooling
until the blackbody is cool.
3.4 Changing Temperature Units
You can select one of three temperature units: Celsius, Fahrenheit, and Kelvin.
To change temperature units:
1. Press the PAGE button until ACCESS appears in the lower display.
2. Press the SCROLL button. LEVEL 1 will appear in the lower display.
3. Press the UP button until CONFIG appears in lower display.
4. Press the UP button until 4 appears and wait.
5. Press PAGE until INSTRUMENT appears.
6. Press the UP button until Units Page appears.
7. Press the SCROLL button. Temp Units appears and °C appears in lower right corner of
the lower display.
Blackbody Setpoint
Blackbody Temperature