M345X12 Blackbody Manual

Operation 15

3.6 Emissivity

When measurements are required at wavelengths other than the one used for factory radiometric
calibration, or if the calibration was done thermometrically (see above), the graph below provides

emissivity vs. wavelength for this model. Integrating over the wavelength region of interest will
provide the average emissivity for your measurement. If the blackbody was calibrated
radiometrically, the apparent emissivity at 8-14um was normalized to 1.0. If the blackbody was
calibrated thermometrically, the target emissivity vs. wavelength can be read directly from the
graph.

Mikron Infrablack 4 - Spectral Em is s ivity Tes t

pe rfor m e d by NIST: March 2005

0.9

0.91

0.92

0.93

0.94

0.95

0.96

0.97

0.98

0.99

1

0

5

10

15

20

25

Wavele ngth (um )

E

m

is

s

iv

it

y

e Mean