Introduction 10

2.3 Tuning Constants

The blackbody source was tuned at the factory. The tuning constants are noted below:

PID 1 (T<103°C)
GAIN (PB) _____
INTEGRAL _______min/repeat
DERIVATIVE______min

PID 2 (T>103°C)
GAIN2 (PB) _____
INT2 ___________
DERV2 _________

2.4 Specifications

Temperature range:

0 to 150.00°C

Accuracy (if thermometric):

±0.05°C

Accuracy (if radiometric):

Refer to NIST traceable calibration report for the
uncertainties of the calibration.

Display Accuracy NIST Cal:

±0.05°C

Temperature resolution:

0.01°C

Stability:

0.1°C per 8 hours period

Source Non-Uniformity:

±0.2°C (excluding 10% Border Area) up to 50°C, 0.6°C @

100°C (4 zones)

Aperture Diameter:

12”x12”

Emissivity
(if thermometric calibration):


0.9756 ±0.0039 @ 8-15 µm 0.9713 ±0.0049 @ 3-5 µm

Emissivity
(if radiometric calibration):


1.0 effective at 8.14 µm

Slew Rate to 1°C stability:

10 minutes maximum for entire range travel

Slew Rate to 0.1°C stability:

15 minutes maximum for entire range travel

Temperature sensor:

Precision platinum RTD 1/3 DIN

Method of control:

Digital PID controller

Operating ambient temperature:

20°C to 30°C

Power requirements:

[ ] 115 VAC ±10% 50/60Hz, 1200 VA maximum
(optional [ ] 230 VAC ±10% 50/60Hz, 1200VA maximum)

Dimensions:

see Appendix