M345X12 Blackbody Manual
Introduction 10
2.3 Tuning Constants
The blackbody source was tuned at the factory. The tuning constants are noted below:
PID 1 (T<103°C)
GAIN (PB) _____
INTEGRAL _______min/repeat
DERIVATIVE______min
PID 2 (T>103°C)
GAIN2 (PB) _____
INT2 ___________
DERV2 _________
2.4 Specifications
Temperature range:
0 to 150.00°C
Accuracy (if thermometric):
±0.05°C
Accuracy (if radiometric):
Refer to NIST traceable calibration report for the
uncertainties of the calibration.
Display Accuracy NIST Cal:
±0.05°C
Temperature resolution:
0.01°C
Stability:
0.1°C per 8 hours period
Source Non-Uniformity:
±0.2°C (excluding 10% Border Area) up to 50°C, 0.6°C @
100°C (4 zones)
Aperture Diameter:
12”x12”
Emissivity
(if thermometric calibration):
0.9756 ±0.0039 @ 8-15 µm 0.9713 ±0.0049 @ 3-5 µm
Emissivity
(if radiometric calibration):
1.0 effective at 8.14 µm
Slew Rate to 1°C stability:
10 minutes maximum for entire range travel
Slew Rate to 0.1°C stability:
15 minutes maximum for entire range travel
Temperature sensor:
Precision platinum RTD 1/3 DIN
Method of control:
Digital PID controller
Operating ambient temperature:
20°C to 30°C
Power requirements:
[ ] 115 VAC ±10% 50/60Hz, 1200 VA maximum
(optional [ ] 230 VAC ±10% 50/60Hz, 1200VA maximum)
Dimensions:
see Appendix