PAGE 12

74845

P2E{} / P2R

= Port 2 serial setup.

INE{} / INR

= Input setup.

TME{} / TMR

= Time.

DTE{} / DTR

= Date.

TIE{} / TIR

= Time/Date Setup

PAE{} / PAR

= Parallel I/O setup

ANE{} / ANR

= Analog setup.

14.3 PRESSURE:

COMMAND:

SUE{}/SUR

DATA FORMAT:

||||||

LENGTH:

Fixed at 13 characters.

WHERE:

d/s

= Display per Second update rate fixed at 1 number code..

ze%

= Zero Aperature % of FS ( Zero %) fixed at 1 number code.

z+%

= Zero Band in % of FS (Zero+/-%) fixed at 1 number code.

azm

= AZM Band in divisions (AZM +/-) fixed at 1 number code.

mot

= Motion Band in divisions (Motion) fixed at 1 number code.

ze set = Auto Tare or Fixed Offset Zero Set) fixed at 1 number code.
min

= Minimum Mode enable (Min) fixed at 1 number code.

Note: The following parameters are not included:

max

= Max Mode enable (Max) fixed at 1 number code

frez

= Freeze mode select.

CODE # / PARAMETER ASSIGNMENT

CODE

#

DSP/

SEC

ZERO

%

ZERO

+/-%

AZM+/-

MOT

ION

ZERO

Set

MIN

0

AUT

O

OFF

FULL

OFF

OFF

OFF

OFF

1

2

0.5

0.05

0.5

1

Auto Tare

ON

2

3

1

0.1

1

3

OffSet

3

5

2

0.2

3

5

4

5

0.5

5

10

5

100

1

10

20

6

1.5

50

14.4 FILTER SETUP:

COMMAND:

FLE/FLR

DATA FORMAT:

LENGTH:

Fixed at 2 characters.

WHERE:

code

= Two (2) digit Filter code # from 00 to 20 per table below.

CODE # / FILTER PARAMETER ASSIGNMENT (FILTR1)

CODE# FILTR 1 CODE# FILTR 1 CODE# FILTR 1 CODE# FILTR 1 CODE# FILTR 1

00

1

05

6

10

12

15

25

20

50

01

2

06

7

11

14

16

30