! Cal Units DATA entry (UUE{} command) and recall ( UUR command)
FORMAT:
LENGTH:
Fixed at two (2) characters.
WHERE:
unit1
: Table III (P=PSI, etc.).
unit2
: Table IV (G=Gage,etc.).
13.0 CALIBRATION:
! General:
The unit must be set for Calibration (Mode Rocker switch 1 or 2 closed).
! COMMANDS:
ZCD
= Perform Zero Cal
RxD
= Perform Linear Reset at point x(x is # from 1 - 9 or letter ‘A’)
LxE{}
= Perform Linear Cal at point x ( x is a # from 1 - 9 or the letter ‘A’).
ZFE{} / ZFR
= Zero Cal correction.
LXE{} / LXR
= Span Cal linear corr. All X's.
LFE{} / LFR
= Span Cal linear corr. All S's.
LCE{} / LCR
= Span Cal linear corr. All C's.
HCE{}
= Perform Hysteresis cal.
HFE{} / HFR
= Hysteresis correction.
! Calibration ‘Perform’ command data entry:
COMMANDS:
LxE{}, HCE{}
DATA FORMAT:
LENGTH:
Fixed at 6 characters.
WHERE:
data:
Six (6) character data field with leading zeros and no
decimal point.
EXAMPLE:
10.00 PSI entered as {001000}
! Calibration ‘Correction’ command data entry and recall:
♦
Zero Cal and Hysteresis correction data;
COMMANDS:
ZFE{}/ZFR, HFE{}/HFR
DATA FORMAT:
LENGTH:
Fixed at 17 characters.
WHERE:
x1, c1 = Eight (8) character data fields
each.
♦
Span Cal Linear corrections (all X’s,S’s, C’s) commands entry and recall data:
COMMANDS:
LXE{}/LXR, LFE{}/LFR, LCE{}/LCR
DATA FORMAT:
LENGTH:
Fixed at 89characters.
WHERE:
x1 - xA: Eight (8) character data field each.
14.0 SETUP DATA:
14.1 GENERAL:
♦
FORMAT:
♦
Setup data is set in a string using the separation character '|' between parameters.
♦
The parameters are listed in the order of the function diagram as they appear in the
manual.
♦
'Arrow' select parameters are represented as a number starting with '0'.
♦
Numeric entered data is directly represented by its data value.
♦
When entering setup data, the full string of parameters need not be entered.
14.2 COMMANDS:
SUE{}/ SUR
= Main Pressure setup.
FLE{} / FLR
= Filter setup.
P1E{} / P1R
= Port 1 serial setup.