Message and corrective actions, Error, Message – Yokogawa DAQWORX User Manual
Page 16

1
IM WX1-01E
.
Message and Corrective Actions
Error
No.
Message
Corrective Actions
E211
Cannot write to file.
Check if the disk capacity is sufficient or if the file systems
is normal.
E212
Cannot read file.
Check if the file exists and is supported by the software or
if the file system is normal.
E213
Cannot open file.
Check if the file exists and is supported by the software or
if the file system is normal
E401
Communication error.
Check if the recorder connected for communication
is powered on and if the cable is properly connected.
Also check the following items according the the
communication type.
• For Ethernet
Check if address settings are correct; the TCP/IP protocol
is installed in Windows; the Ethernet card is properly
installed.
• For RS-232 and RS-422-A
Check if the baud rate settings match; the port (COM1 to
COM9) settings match, the address settings are correct
(RS-422-A); the serial port of the PC is active and the
appropriate cable is being used.
E402
Communication timeout.
-
E403
Cannot open a communication port.
Same as E401.
E501
Invalid license number. Please reinstall the software.
Install the software again.
E1010
Execution of a process failed.
Check whether an executable function exists, or whether
its files are damaged. If this error appears frequently,
reinstall the software.
E1011
Execution of a service failed.
Check whether an executable function exists, or whether
its files are damaged. If this error appears frequently,
reinstall the software.
Message
No.
Mesage
M1201
Model determination was successful.
M1210