Error Messages and Correctve Actons
A message (such as an error message) may appear during operation. The following
describes the meanings of the messages and their corrective actions.
Error
Code
Message
Correctve Actons
E211
Cannot write to the file.
Confirm that the disc capacity and file system are correct.
E212
Cannot read file.
Confirm that a file exists, that the software supports it, and whether the file system
is correct.
E213
Cannot open the file.
Confirm that a file exists, that the software supports it, and whether the file system
is correct.
E401
Communication error.
Check whether the power to the connected measuring instruments with which
you wish to open communications is ON, and whether the cables are connected
correctly. Also check the following.
• For Ethernet
Is the address correct, is the TCP/IP protocol installed in Windows, and is the
Ethernet card installed correctly? Is the address correct, is the TCP/IP protocol
installed in Windows, and is the Ethernet card installed correctly?
• For RS-232 and RS-422A
Are the baud rate, port numbers (COM1–COM9), and addresses (RS422)
correct? Is the PC serial port valid? Is the cable type correct?
E402
Communication timeout.
–
E403
Communication open error.
Check whether the power to the connected measuring instruments with which
you wish to open communications is ON, and whether the cables are connected
correctly. Also check the following.
• For Ethernet
Is the address correct, is the TCP/IP protocol installed in Windows, and is the
Ethernet card installed correctly?
• For RS-232 and RS-422A
Are the baud rate, port numbers (COM1–COM9), and addresses (RS422)
correct? Is the PC serial port valid? Is the cable type correct?
E501
Invalid license number.
With the Gate series, the number of licenses purchased is the limit for the number
of programs run at the same time.
E1001
Login failed.
Check the user name and password on the µR10000.
E1002
Invalid password.
Check the user name and password on the µR10000.
E1003
Invalid password.
Check the user name and password on the µR10000.
E1010
A user with the same name is
already logged in.
Check the user name and password on the µR10000.
Message
Code
Message
M1201
Model determination completed successfully.
M1210
Save the setting changes and try again.