Precautions to be taken when loading setup data – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 8

8
IM 04L31A01-05E
Saving the Settings
1. Press the ESC key.
The display returns to the basic setting menu.
2. Press the
[End] soft key
.
A window appears for you to confirm the saving of the new settings.
3. Select [Yes] and press the
DISP/ENTER key
.
Precautions to Be Taken When Loading Setup Data
As a general rule, load setup data of the same firmware version. If you must load setup
data of different versions, note the following points.
•
When loading setup data that has been saved on version 3.01 using version 3.02
Event output settings (time event, PV event, and program pattern end signal settings)
in Setting mode (Control) are not loaded.
•
When loading setup data that has been saved on version 3.02 using version 3.01
• Event output setting (common or separate) in Basic Control Settings is not loaded.
• Event output settings (time event, PV event, and program pattern end signal
settings) in Setting mode (Control) are not loaded.
• Setup data is not loaded when the displayed language is set to Chinese.
Operations on the CX1000/CX2000