3.3 Error Messages and Corrective Actions
Errors That Occur during Authentication
Code
Message
Explanation/Corrective Action
E006
Incorrect input character string.
Check that the host principal, authentication key
password, and realm name settings on the DX are
correct.
E085
The login password is incorrect.
Enter the correct password.
E110
This user name is not registered.
The specified user is not registered on the DX. The user
account is not registered on the server.
E114
This user name is invalid.
The account has been invalidated on the server.
The account has been invalidated on the DX.
E117
This password is not effective.
On the DX, because the wrong password has been
entered consecutively for more than the permissible
number of times, this user is invalid.
E764
Not supported by this machine.
Not supported by the DX.
E765
Preauthentication failed.
Enter the correct password. Also, make sure that the
times on the DX and the server match.
E766
The encryption type is not supported by this machine.
The DX does not support the encryption type, or the
encryption type settings on the DX and the server are
different. Use the same encryption method on the DX and
the server.
E767
Failed to receive authentication from KDC server.
Check the DX and server settings. Also, make sure that
the times on the DX and the server match.
E768
Change the password.
The password has expired. Change the password of the
user account that is registered on the server.
E769
The time difference with the KDC server exceeds the
limit.
There is a time difference of 5 minutes or more between
the DX and the server. Set the DX time to match the time
on the server.
E771
The host principal is not registered.
The host account is not registered on the server.
E772
The host principal is invalid
Check the host account that is registered on the server.
E773
The host password is incorrect.
Make sure that the DX authentication-key password and
the server's host-account password match.
E774
Preauthentication failed.
An internal error occurred during preauthentication.
Disable the server's preauthentication function.
E775
The realm is incorrect.
Make sure that the realm name setting on the DX is
correct.
Errors That Occur during Communication
Code
Message
Explanation/Corrective Action
E260
IP address is not set or ethernet function is not
available.
The server address has not been specified.The Ethernet
is not functioning.
E266
Ethernet cable is not connected.
Make sure that an Ethernet cable is connected.
E760
Cannot find KDC server.
The KDC server cannot be found in the same domain.
E761
KDC server connection error.
An error occurred while the DX was connecting to the
KDC server. Make sure that the network connection is not
broken.