1.5 Audit Trail Function

This function records histories of the operations. It saves operation logs and change
settings logs, and it saves setup files when the settings have changed. You do not need
to perform any special settings to use this function.
The figure below indicates what items are recorded to the operation log and the change
settings log.

Memory Stop

Setting change

Memory Start

Setting change

Memory Stop

Operation Log

Setup file

Display data files or

event data files

Setup file

Change

Settings Log

Setting

Change

Audit Trail Function

Change Settings

Log

Operation

Log

Record range

Record range

Information That Is Saved to Measured Data Files

When measured data files (display data or event data files) are saved, in addition to the
measured data, a setup file and operation log are also saved.

Setup File

A file that contains the settings at the time recording started (memory start). If the
settings are changed during recording (memory sampling), you can view the changes in
the operation log.

Operation Log

A history of the operations.